Multitron - Operating Manual
Options
Page
36
of
158
20 August 2020
The device can be equipped with an internal cooling unit. The cooling
unit can either be placed in a bracket on the incubation shaker or in
the high base (900 W version only).
The cooling unit is available in two different versions:
Version with 900 W cooling capacity (one compressor)
Version with 1800 W cooling capacity (two compressors)
The version with 1800 W cooling capacity is used when very low
temperatures have to be reached. As two independent compressors
are installed in this version, it is also ensured that cooling can con-
tinue if one compressor fails (redundancy).
The cooling unit is an autonomous system and has its own power
supply and mains switch. This is located on the right-hand side of the
bracket or base. The cooling unit is a 2-circuit system with a coolant
tank. Prior to commissioning, it must be filled with coolant by the
manufacturer’s qualified expert personnel or persons authorised by
the manufacturer.
The heat exchanger is located behind the rear wall in the incubation
chamber.