Multitron - Operating Manual
Operation
Page
80
of
158
20 August 2020
Tray cannot be inserted
If you are unable to insert the tray correctly, perform the following
checks:
Remove the tray and check it is not twisted or bent.
Check the incubation chamber for foreign matter and other ob-
jects.
If it is still not possible to insert the tray correctly, contact your
local INFORS HT representative.
7.2.3
Fitting the Holders
ATTENTION
The threaded holes of the tray can be damaged if the fastening
screws are not screwed straight into the thread:
–
Place the screws vertically on the drill holes.
–
Ensure they can be tightened easily.
ATTENTION
All holders are supplied with pre-fitted screws. If a test tube holder
or a box for microtiter plates is placed on the Sticky Stuff adhesive
matting, there is a risk that the adhesive matting is damaged by
protruding screws.
Remove the screws before you place test tube holders or boxes for
microtiter plates on the Sticky Stuff adhesive matting.
Fitting a holder
Clamps, test tubes and boxes for microtiter plates are fixed to the
tray using screws. The device is delivered with pre-fitted screws (2).
The screws are each secured with one flat gasket at the bottom (1) so
that they cannot fall out. Only use the screws provided or screws of
the same size for assembly.
To fit a holder, proceed as follows:
1
2