Multitron - Operating Manual
Installation
20 August 2020
Page
73
of
158
To achieve a stable climate in the incubation chamber and to be able
to control the climate accurately, the following installation locations
must be avoided:
Poorly ventilated recesses or in the exhaust heat air stream of air
conditioners or other sources of heat
Directly underneath, next to or in the air stream of the air condi-
tioners or other devices that generate strong air flows (e.g. cir-
culation fans in clean rooms)
Next to devices with strong heat radiation, such as heaters or
autoclaves, or in their exhaust heat air stream
Next to devices for refrigeration, such as ultra-low temperature
freezers or refrigerated centrifuges, or in their exhaust heat air
stream
INFORMATION
In order to achieve the desired temperature in the incubation
chamber, the ambient temperature directly at the device is deci-
sive. This can be considerably higher than the temperature at other
positions in the room due to waste heat from the device or other
devices with strong heat radiation in the immediate vicinity!
6.2
Requirements for the Mains Connection
To avoid dangers due to electrical current, the in-house mains con-
nection must meet the following requirements:
Single-phase, constant power supply
In the building, the power supply must be protected using a
fault current protection switch (RCD – Residual Current Device).
In addition to that, the following requirements must be met:
Make sure that the voltage values of the device match those of
the local power supply. Note the description on the identifica-
tion plate.
Only use the power cable provided. If the power cable becomes
defective, replace it with a power cable of the same type.
Ensure that the mains connection is accessible at all times.