VOLTage
— This element provides the voltage
measurement or the programmed voltage source
reading. If sourcing voltage and measuring voltage,
this element will provide the voltage measurement
(measure reading takes priority over source reading). If
voltage is not sourced or measured, the NAN (not a
number) value of +9.91e37 is used.
CURRent
— This element provides the current
measurement or the programmed current source
reading. If sourcing current and measuring current,
this element will provide the current measurement
(measure reading takes priority over source reading). If
current is not sourced or measured, the NAN (not a
number) value of +9.91e37 is used.
RESistance
— This element provides the resistor
measurement. If resistor is not measured, the NAN
(not a number) value of +9.91e37 is used.
TIME
— A timestamp is available to reference each
group of readings to a point in time. The relative
timestamp operates as a timer that starts at zero
seconds when the instrument is turned on or when the
relative timestamp is reset (:SYSTem:TIME:RESet). The
timestamp for each reading sent over the bus is
referenced, in seconds, to the start time. After
99,999.999 seconds, the timer resets to zero and starts
over.
Timestamp values are approximate.
Timestamp is also available for buffer readings.
Timestamp can be referenced to the first reading stored
in the buffer (absolute format) which is timestamped at
0 seconds, or can provide the time between each
reading (delta format). The :TRACe:TSTamp:FORMat
Command is used to select the absolute or delta
format.