For example, if voltage and current measurements
were stored in the buffer, then the selected statistic
operation will be performed on both readings.
Statistics for multiple measurement functions are
returned in the following order:
voltage statistic, current statistic, resistor statistic.
Statistic operations are not performed on TIME and
STATus data elements that are stored in the buffer.
If the buffer is configured to store the result of CALC1
or CALC2 (:TRACe:FEED CALC1 or CALC2), only one
result will be returned by this query Command.
NOTE:
1. If there is no data in the buffer, error -230, “Data
corrupt or stale,” will be generated.
2. If there are a lot of readings stored in the buffer,
some statistic operations may take too long and cause
a bus time-out error. To avoid this, send
the :CALA3:DATA? Command and then wait for the
MAV (message available) bit in the Status Byte
Register to set before addressing the GSM to talk.
Example
:CALCulate3:DATA?