GSM-20H10 User Manual
293
Command
[:SENSe[1]]:FUNCtion:OFF?
Function
Query functions that are disabled
Example
:SENSe1:FUNCtion:OFF?
Command
[:SENSe[1]]:FUNCtion:OFF:ALL
Function
This Command is used to disable all measurement
functions. When enabled (:ON:ALL), amps, volts,
and ohms measurements will be performed
simultaneously if concurrent measurements are
enabled.
If concurrent measurements are disabled,
only the ohms function will be enabled.
The :OFF:ALL Command disables all measurements.
Example
:SENSe1:FUNCtion:OFF:ALL
Command
[:SENSe[1]]:FUNCtion[:ON]:COUNt?
Function
This query Command is used to determine the
number of functions that are enabled.
When :ON:COUNt? is sent, the response message
will indicate the number of functions that are
enabled.
Example
:SENSe1:FUNCtion:ON:COUNt?
Command
[:SENSe[1]]:FUNCtion:OFF:COUNt?
Function
This query Command is used to determine the
number of functions that are disabled.
When :OFF:COUNt? is sent, the response message
will indicate the number of functions that are
disabled.
Example
:SENSe1:FUNCtion:OFF:COUNt?
Command
[:SENSe[1]]:FUNCtion:STATe? <name>