![GW Instek GSM-20H10 User Manual Download Page 269](http://html1.mh-extra.com/html/gw-instek/gsm-20h10/gsm-20h10_user-manual_2280064269.webp)
GSM-20H10 User Manual
267
:SPAN? MINimum Query lowest allowable level
of V-Source
:SPAN? MAXimum Query highest allowable level
of V-Source
Example
:SOURce:VOLTage:SPAN?
Command
:SOURce[1]:CURRent:STEP <n>
Function
This Command is used to specify a step size for a
linear sweep. When the sweep is started, the source
level changes in equal steps from the start level to the
stop level. A measurement is performed at each
source step (including the start and stop levels).
<n> = -2.1 to 2.1 Set I-Source level (amps)
DEFault 0A
MINimum -2.1A
MAXimum +2.1A
This Command cannot be used for a logarithmic
sweep. Use the :POINTs Command to set the source-
measure points for a log sweep.
To avoid a setting conflicts error, make sure the step
size is greater than the start value and less than the
stop value.
The number of source-measure points in a linear
sweep can be calculated as follows:
Points=[(Stop-Start)/Step]+1 or
Points=(Span/Step)+1
An alternate way to set the source-measure points in
a linear sweep is to simply specify the number of
source-measure points in the sweep using
the :POINts Command.