GSM-20H10 User Manual
343
Example
:ARM:OUTPut TENTer
Command
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OUTP
ut?
Function
Query output trigger event(s)
Example
:ARM:OUTPut?
Command
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut
<event list>
Function
This Command is used to specify when trigger pulses
occur on the specified output trigger line of the
Trigger Link.
Trigger Layer Triggers — You can specify from one
to all three events.
Each event in the list must be separated by a comma
(,).
The SOURce, DELay, and MEASure events refer to
the Source-DelayMeasure (SDM) cycle. This is the
Device Action in the Trigger Model.
With SOURce
specified, an output trigger occurs after the source is
set. With DELay specified, an output trigger occurs
after the delay period. With MEASure specified, an
output trigger occurs after the measurement.
Trigger Layer Triggers
<event list>: SOURce Output trigger after
source level is set
DELay Output trigger after delay
period