OVERVIEW
18
Description The display is used primarily to program source and
compliance values and display the real measured
readings, The area is divided as follows.
①
Status and Error messages
Status and error messages are displayed
momentarily, which are located on the top of the
display, indicate various states of operation. The
meaning of each status information is described
separately in each function introduction chapter.
The error message contains a variety of error
types, which can be read by instructions under
remote control. Typical error messages
description and types are listed in page 358.
②
Reading Data/Format
The readback value is displayed with a resolution
of up to 6
1/2
digits. The number of display digits
can be changed by directly setting Digits or
modifying the refresh Speed. Take the highest
resolution of the readback voltage as an example:
The readback display area is divided into 3 parts:
1: sign, 2: value, 3: units.
˙
The measured values of voltage, current and
resistor can be positive or negative.
˙
The digital display has a total of 7 digits at
most, and the range is determined by the
integer digits combined with the unit.
When measuring V:
˙
3 integer digits for 200V or 200mV range
˙
2 integer digits for 20V range
˙
1 integer digits for 2V range