GSM-20H10 User Manual
69
Generally speaking, the fastest sampling speed (Fast:
0.01PLC) will result in an increase in reading noise
and a decrease in the number of available digits. The
slowest sampling speed (High: 10PLC) provides the
best accuracy and noise suppression. The middle
settings are compromise between speed and noise.
The default power on speed is Normal (1PLC). Other
is a custom option.
The method of setting the sampling speed: press the
arrow keys to move the cursor to the Speed setting
box.
·
Fast- Setting the sampling speed to 0.01 PLC, and
the Digits option is automatically set to 3
1/2
digits.
·
Medium-Setting the measurement speed to 0.10
PLC, and the Digits option is automatically set to
4
1/2
digits.
·
Normal-Setting the measurement speed to 1.00
PLC, and the Digits option is automatically set to
5
1/2
digits.
·
High-Setting the measurement speed to
10.00PLC, and the Digits option is automatically
set to 6
1/2
digits.
·
Other-Used to set the measurement speed to any
PLC between 0.01 PLC and 10 PLC. When Other
is selected, the cursor automatically jumps to the
PLC quantity window, press Enter to write the
customized value, and then Enter to confirm.
When using this option to set the speed, the
Digits option will not be changed.
NOTE: After setting the measurement speed,
you can use the Digits option to change the number
of displayed digits.
*PLC stands for Power Line Cycle.
NOTE: Changing the Speed will modify the