COMMAND SYNTAX
266
Command :SOURce[1]:VOLTage:SPAN <n>
Function
A sweep can be configured by specifying center and
span parameters.
By specifying a center point, you can sweep through
the operating point of a device. The span determines
the sweep width with the operating point at the
center of the sweep.
For example, assume you are testing a device that
operates at 10V, and you want to sweep from 8 to 12
volts. To do this, you would specify the center to be
10V and the span to be 4 volts (12 - 8).
Use the :STEP or :POINts Command to specify the
number of sourcemeasure points in the sweep.
:CENTer and :SPAN are coupled to :STARt
and :STOP. Thus, when center and span values are
changed, the values for start and stop are affected as
follows:
Start=Center-(Span/2)
Stop=(Span/2)
<n> =-420 to 420 Set V-Source level (volts)
DEFault 0V
MINimum -420V
MAXimum +420V
Example
:SOURce:VOLTage:SPAN 0
Command
:SOURce[1]:VOLTage:SPAN?
Function
:SPAN? Query span for sweep of
V-Source
:SPAN? DEFault Query *RST default level of V-
Source