GSM-20H10 User Manual
245
multiple source/measure functions to be used in a
sequence.
<name> = VOLTage Select V-Source mode
CURRent Select I-Source mode
MEMory Select memory mode
Example
:SOURce:FUNCtion VOLTage
Command
:SOURce[1]:FUNCtion[:MODE]?
Function
Query the type of selected source.
Example
:SOURce:FUNCtion?
Command
:SOURce[1]:CURRent:MODE <name>
Function
This Command is used to select the DC sourcing
mode for the I-source. The three modes are explained
as follows:
FIXed — In this DC sourcing mode, the specified
source will output a fixed level. Use the :RANGe
and :AMPLitude Commands to specify the fixed
source level.
LIST — In this mode, the source will output levels
that are specified in a list.
SWEep — In this mode, the source will perform a
voltage, current or memory sweep.
<name> = FIXed Select fixed sourcing mode
LIST Select list sourcing mode
SWEep Select sequence sourcing mode
Example
:SOURce:CURRent:MODE FIXed
Command
:SOURce[1]:CURRent:MODE?
Function
Query DC sourcing mode.