GSM-20H10 User Manual
113
Press the F4 (Pass) key for the related actions under
PASS. Related instructions can refer to page 123.
Pass pattern
Used to define the digital output bit value. When the
Digout size is 3 digits, the value digits are 0-7, and
when the Digout size is 4 digits, the value digits are 0-
15. When Mode is set as Grading, output the Pass
pattern value in case that all limit tests pass. When
Mode is set to Sorting, it is the pass pattern value of
Limit 1 (Compliance Test) when SW-Limits are
Disable.
Setting: Move the cursor to the value of the
corresponding item (it becomes a red letter on a gray
background), press Enter, and it is in the
programming state. Operate the direction keys and
numeric keys to input the desired value.
Source
memory
location
It is used with SRC-MEM sequence to select the next
memory location point in the sequence in the PASS
condition. If selecting Next, the next location of the
present location in the sequence list is selected. If
selecting Location, you can jump to any location from
the present location in the sequence list (LOCATION 1
to LOCATION 100) by setting the location number.
Setting: Move the cursor to the corresponding item (it
turns into red on the gray background), press Enter,
when the up and down arrow signs appear, operating
the up and down direction buttons to select the
required option (Next or Location), and then press
Enter to confirm.
Location
Specify the branching Location in the sequence list,
the range is 1-100. Only need to be set when Source
memory location is selected as Location.
Setting: Move the cursor to the value of the
corresponding item (it becomes a red letter on a gray