Example
:SENSe1:Resistance:MODE?
Command
[:SENSe[1]]:RESistance:OCOMpensated <b>
Function
This Command is used to enable or disable offset-
compensated ohms. When using the auto ohms
measurement mode, the current source level is
automatically set. When using the manual ohms
measurement mode, you must set the source (V or I)
output level.
Offset-compensated ohms will disable when
the :MEASure? Command (for the resistance
function) or the :CONFigure:RESistance Command is
sent.
NOTE:
1. You cannot select a current measurement range if
sourcing current. Conversely, you cannot select a
voltage measurement range if sourcing voltage.
Also, autorange cannot be enabled for those source-
measure configurations. The programmed source
range determines measurement range.
2. You cannot select an ohms measurement range if
in manual ohms (you must be in auto ohms).
3. The highest current measurement range that can
be selected is limited by the current compliance
range. For example, if current compliance is set for
50mA (100mA range), then the highest available
current measurement range is 100mA. Similarly, the
highest voltage measurement range is limited by
the voltage compliance range.
4. Measurement range limitations: With the 200V
source range selected, the highest current
measurement range is 100mA. With the 1A source
range selected, the highest voltage measurement
range is 20V.