![GW Instek GSM-20H10 User Manual Download Page 143](http://html1.mh-extra.com/html/gw-instek/gsm-20h10/gsm-20h10_user-manual_2280064143.webp)
GSM-20H10 User Manual
141
TRIG in
Source
Select the trigger source of the Trigger layer, which
can be set as Immediate or Trigger link.
Immediate:
Indicates that the event is triggered
immediately. When this option is
selected, source event detector, delayed
event detector and measurement event
detector will run immediately. The
Trigger layer executes the source, delay,
and measurement actions in sequence.
Trigger link:
If you select this option, you need to set
these 5 items of Link/Detect bypass/
Events source /Events delay/ Events
MEAS.
Link
To select the input line of the trigger signal, one of
the 4 lines (#1, #2, #3, #4) can be selected.
Detect
bypass
Used to set whether to bypass the source event
detector
, you can choose Once or Never.
Once:
Indicates that the operation will bypass the
event detector.
Never:
Indicates that the event detector needs to
wait for an input trigger signal before
performing subsequent operations.
Events
source
Enable (On) or disable (Off)
trigger-in source.
On:
Indicate that operation will wait at that event for