![GW Instek GSM-20H10 User Manual Download Page 113](http://html1.mh-extra.com/html/gw-instek/gsm-20h10/gsm-20h10_user-manual_2280064113.webp)
GSM-20H10 User Manual
111
Operate the direction keys or numeric keys to input
the desired value.
SW-Limits
Press the F3 (SW-Limits) key to control the
enabling of LIM2, LIM3, LIM5-LIM12, High/Low
limit range and fail pattern value. The specific
settings are shown in the figure below:
Disable/
Enable
Set the function of LIM2, LIM3, LIM5-LIM12 to
Enable.
Setting: Move the cursor to the corresponding
Disable or Enable item (turned into red on a gray
background), press Enter, and when the up and
down arrow signs appear, operating the direction
keys to select the required option (Disable or
Enable), and then press Enter to confirm.
Low
Set the low value of the limit range of LIM2, LIM3,
LIM5 -LIM12.
Setting: Move the cursor to the value of the
corresponding item (it becomes a red letter on a
gray background), press Enter, and it is in the
programming state. Operating the up and down
direction keys to select the required magnification,
using the left and right direction keys and the