COMMAND SYNTAX
318
Command
:SYSTem:POSetup?
Function
Query power-on setup
Example
:SYSTem:POSetup?
Command
:SYSTem:RSENse <b>
Function
This Command is used to enable or disable remote
sensing. When using remote sensing, 4-wire
connections to the DUT are required.
Sourcing voltage—With remote sensing enabled, the
output voltage is sensed (measured) at the DUT. If
the sensed voltage is lower than the programmed
amplitude, then the V-Source increases the voltage
until the sensed voltage is the same as the
programmed amplitude. This compensates for IR
drop in the OUTPUT test leads.
With remote sensing disabled, the output voltage is
sensed at the output connectors.
Measuring voltage — With remote sensing enabled,
voltage measurements are made at DUT. This
eliminates any voltage drops that may be in the test
leads between the GSM and the DUT.
With remote sensing disabled, voltage measurements
are performed at the output connectors of the
instrument.
Measuring resistor — With remote sensing enabled,
4-wire resistor measurements can be made.
<b> = 0 or OFF Disable remote sensing
1 or ON Enable remote sensing
Example
:SYSTem:RSENse 0
Command
:SYSTem:RSENse?
Function
Query state of remote sensing.