Example
:SYSTem:RSENse?
Command :SYSTem:GUARd <name>
Function
This Command is used to select the guard mode.
OHMS guard is a lowimpedance guard drive used for
in-circuit resistor measurements.
Ohms guard is not available 1A range (source or
measure). CABLE guard provides a highimpedance
guard drive that is used to eliminate leakage currents
in cabling and test fixtures.
When performing 6-wire ohms guard measurements,
use the GUARD out-put state. The OUTPut
[1]:SMODe GUARd Command is used to select the
GUARD output-off state.
<name> = OHMS Ohms guard mode
CABLe Cable guard mode
Example
:SYSTem:GUARd OHMS
Command
:SYSTem:GUARd?
Function
Query guard mode.
Example
:SYSTem:GUARd?
Command
:SYSTem:MEMory:INITialize
Function
When this Command is used, the following actions to
initialize battery backed RAM occur:
• TRACe (data store) data is lost, buffer size is reset
to 100, and timestamp is set to the absolute format.
• SOURce1:LIST:CURR and VOLT are reset to 0A
and 0V, respectively.
• Deletes all user-defined math expressions.