OVERVIEW
22
Maximum compliance value
The following table summarizes the maximum
compliance values of the measuring range:
Measure range
Max. Cmpl value
Voltage
200mV
±210mV
2V
±2.1V
20V
±21V
200V
±210V
Current
1uA
±1.05uA
10uA
±10.5uA
100uA
±105uA
1mA
±1.05mA
10mA
±10.5mA
100mA
±105mA
1A
±1.05A
The Cmpl setting of voltage or current contains
three aspects, as shown in the figure:
a
,
The choice of Vcmpl or Icmpl is determined
by source. When Vsrc is selected, Icmpl is
automatically selected. Similarly, when Isrc is
selected, Vcmpl is automatically selected.
b
,
Vcmpl or Icmpl range setting: The Edit/Lock
button on the operation panel can alternately
select SRC, Cmpl or parameter area. When
the Cmpl value area is marked, operate the
up and down direction keys and observe the
position and unit of the decimal point to
understand the corresponding range.
c
,
Vcmpl or Icmpl value setting: when the Cmpl
value area is marked, input the required
value from the numeric keyboard according
to the high and low digits, or use the left and
right arrow keys to select the digit to input
the corresponding value.