GSM-20H10 User Manual
149
Reset time
stamp
Used to reset the time stamp when exiting the idle
state. There are two options, Yes or No.
Yes:
In the trigger mode, when exiting the idle
mode, the time stamp is allowed to be reset
automatically.
No:
In the trigger mode, when exiting the idle
mode, the time stamp is not allowed to be
reset automatically.
Numbers Used to select Engineering units or scientific notation
to display the measurement reading. The
measurement reading can be fixed or floating-point
format.
Beeper
Used to select Enable or Disable touch-tone, the
factory default is Enable. For Beeper application,
please refer to the instructions on page 320.
Memory
save
Used for source memory sequence setup
configurations. Up to 100 setups can be stored. Enter
the number (1-100) in the option box and press Enter
to save the present setup in the corresponding
location in the memory.
The stored setups can be recalled directly (operate by
Memory restore) or recall multiple setups
continuously (see the SRC-MEM chapter on page 123
for details)
Memory
restore
It is used to recall the setups in the memory. Up to 100
setups can be recalled. Enter the number (1-100) in the
option box and press Enter to recall the setup
information stored in the corresponding location in
the memory to the Source interface.
Output
There are several parameters related with Output
required to be set here: Off state, Auto off, Enable,
Front/Rear Output.
Off state: Used to select the off state of the Output
terminal. There are four options: High impedance,
Normal, Zero and Guard.