BASIC OPERATION
72
NOTE:
a
Relative Value is valid to all the ranges. For
example, if the Relative Value is set to 5V in the
20V range, when the range is changed to 2V or
200V, the Relative Value is still 5V.
b
When the programmed Relative Value exceeds
the slected range, it will not cause overflow and
will not increase the maximum allowable input
for the range. On the 20V range, for >21.1V
input, the GSM will still overflow.
c
When Relative is set as Enable, the display area
will display REL. If toggling the V/I
measurement, Relative option will
automatically change to Disable.
Enter a Relative Value manual:
1)
Press the arrow keys to move the cursor to the
Relative Value setting box, select Enable, and
then press Enter.
2)
Press the arrow keys to move the cursor to the
Value setting box, set the required value, and
then press Enter.
Other
Settings
Long press the Source key to open the following
interface which are used to set the parameters of OVP,
Auto delay, Vsrc trig control, and Isrc trig control.