COMMAND SYNTAX
254
Command :SOURce[1]:VOLTage:PROTection[:LEVel] <n>
Function
This Command is used to set the Over Voltage
Protection (OVP) limit for the V-Source. The V-Source
output will not exceed the selected limit. When Vsrc
exceed the value of OVP, voltage of the input/output
port will be the value of OVP.
The OVP limit is also enforced when in the I-Source
Mode.
The limit parameter values are magnitudes and are in
effect for both positive and negative output voltage.
You can express the limit as a positive or negative
value.
<n> = -210 to 210 Specify V-Source limit
NONE Disable OVP function
DEFault Set limit to 210V
MINimum Set limit to 20V
MAXimum Set limit to 210V
Example
:SOURce:VOLTage:PROTection 20
Command
:SOURce[1]:VOLTage:PROTection[:LEVel]?
Function
[:LEVel]? Query voltage protection
limit level
[:LEVel]? DEFault Query *RST default limit
[:LEVel]? MINimum Query lowest allowable limit
[:LEVel]? MAXimum Query highest allowable limit
Example
:SOURce:VOLTage:PROTection?
Command
:SOURce[1]:DELay <n>