GETTING STARTED
34
Manually set from the front panel: System->Control->
Rear, or set by the Command: :ROUTe:TERMinals.
NOTE: After change the front and rear terminals,
the output will be turned off.
Connect To Load
Description The maximum allowable voltage
differentials between terminals
are labeled on the Front/Rear
panel, as shown in the figure
(rear panel).
WARNING: To prevent electric shock and
damages to the GSM-20H10, DO NOT exceed the
maximum allowable voltage differentials shown in
terminals. The front and rear terminals of the GSM
are rated for connection to circuits rated Installation
Category I only. Do not connect the GSM terminals to
CAT II, CAT III, or CAT IV circuits. Connections of
the GSM Input/Output terminals to circuits higher
than CAT I can cause damage to the equipment or
expose the operator to hazardous voltages.
To prevent electric shock and damage to the GSM,
external common mode voltage must be limited within
250VDC, 1.05A maximum.