GSM-20H10 User Manual
243
Source Commands
Command
:SOURce[1]:CLEar[:IMMediate]
Function
This Command is used to turn off the source output.
The output will turn off after all programmed source-
measure operations are completed and the
instrument returns to the idle state.
NOTE : if Auto off is enabled, the source output
will automatically turn off.
Example
:SOURce:CLEar
Command
:SOURce[1]:CLEar:AUTO <b>
Function
This Command is used to control auto output-off for
the source. With auto output-off enabled, an :INITiate
(or :READ? or MEASure?) will start source-measure
operation. The output will turn on at the beginning of
each SDM (source-delay-measure) cycle and turn off
after each measurement is completed.
With auto output-off disabled, the source output
must be on before an :INITiate or :READ? can be
used to start source-measure operation.
The :MEASure? Command will automatically turn on
the source output.
Once operation is started, the source output will stay
on even after the instrument returns to the idle state.
Auto output-off disabled is the *RST
and :SYSTem:PRESet default.
WARNING: With auto output-off disabled, the
source output will remain on after all programmed
source-measure operations are completed. Beware
of hazardous voltage that may be present on the
output terminals.