GSM-20H10 User Manual
95
d
Press the 0/M key to switch to the calculation
Measure interface. The VarAlpha symbol
appears in the display area, indicating that the
VarAlpha to be calculated.
e
Finally, turn on the Output key, and the
measurment reading of VarAlpha will be
displayed.
NOTE: When programming the value of I1 and
I2, the up and down direction keys can switch the
current range, and the value should be set in the
most suitable range according to the range of the
resistor to be measured.
For example, when the GSM is connected to a 1MΩ
pure resistor, I1 is set to 0.01uA, and I2 is set to
1.05uA, then the parameter settings and
measurement reading are shown in the following
figure:
DEV
Used to measure the percentage deviation between
the measurement reading and the reference value set
by the user.