GSM-20H10 User Manual
153
connect a spark discharge tube between
INPUT HI and LO.
Auto off:
Used to Enable or Disable "output
automatic off" function.
Enable: Output will be turned off after the
end of the measurement phase of
each SDM cycle. Output will be
turned on again at the beginning
of the next SDM cycle.
Disable: Output remains on as long as the
GSM operates in the trigger mode
(the ARM symbol is displayed).
NOTE: In the case of Enable,
pressing the Output key will turn off
the Auto off function (it becomes the
Disable state).
Enable:
Enable or Disable the output enable function
(/OE line). Enable or Disable Output when
the GSM is connected to the DUT or test
fixture equipped with a switch controlled by
lid. For details, see External I/O port
operation on page 115 of the Limit chapter.
Enable: Turn on the output enable function.
The output of the GSM is controlled
by the input signal of /OE line.
Disable: Turn off the output enable function.
Rear:
Used to select the rear-panel output.
Enable: Input/Output is performed from the
rear panel.
Disable: Input/Output is performed from the
front panel.
Filter
Used to turn on (Enable) or turn off (Disable) the
output filter.
Average mode:
When Filter is set as Enable, you need
to set this option. It is used to select