GSM-20H10 User Manual
311
Example
:SENSe1:AVERage:TCONtrol?
Command
[:SENSe[1]]:AVERage:COUNt <n>
Function
These Commands are used to specify the filter count.
In general, the filter count is the number of readings
that are acquired and stored in the filter buffer for
the averaging calculation. The larger the filter count,
the more filtering that is performed.
<n> = 1 to 100 Specify filter count
DEFault 10
MINimum 1
MAXimum 100
Example
:SENSe1:AVERage:COUNt 1
Command
[:SENSe[1]]:AVERage:COUNt?
Function
:COUNt? Query filter count.
:COUNt? DEFault Query the *RST default filter
count.
:COUNt? MINimum Query the lowest allowable
filter count.
:COUNt? MAXimum Query the largest allowable
filter count.
Example
:SENSe1:AVERage:COUNt?
Command
[:SENSe[1]]:AVERage[:STATe] <b>
Function
This Command is used to enable or disable the digital
averaging filter. When enabled, voltage, current, and
resistor readings are filtered according to how the
filter is configured.