COMMAND SYNTAX
332
Function
When this action command is sent, any pending
(latched) input triggers are cleared immediately.
When the GSM is being triggered by another
instrument, it may inadvertently receive and latch
input triggers that do not get executed. These
pending triggers could adversely affect subsequent
operation.
When using external triggering, it is recommended
that TRIGger:CLEar be sent after sending the ABORt
command and at the beginning of a program before
sending an initiate Command.
Example
:TRIGger:CLEar
Command
:INITiate[:IMMediate]
Function
This Command is used to initiate source-measure
operation by taking the GSM out of idle. The :READ?
and :MEASure? Commands also perform an
initiation.
NOTE that if auto output-off is disabled
(SOURce1:CLEar:AUTO OFF), the source output
must first be turned on before an initiation can be
performed. The :MEASure? Command automatically
turns the output source on before performing the
initiation.
Example
:INITiate
Command
:ARM[:SEQuence[1]][LAYer[1]]:COUNt <n>
Function
This Command is used to specify how many times an
operation is performed in the arm layer of the trigger
model.
<n> = 1 to 2500 Specify arm count
DEFault Sets arm count to 1
MINimum Sets arm count to 1