COMMAND SYNTAX
240
Example
:OUTPut:ENABle 0
Command
:OUTPut[1]:ENABle:STATe?
Function
Query state of output enable line.
Example
:OUTPut:ENABle:STATe?
Command
:OUTPut[1]:ENABle:TRIPped?
Function
This query Command is used to determine output
enable has been tripped. The tripped condition (“1”)
means that the source meter can output (output
enable line at logic low level). A “0” will be returned
if the source meter cannot output (output enable line
at logic high level).
Example
:OUTPut:ENABle:TRIPped?
Command
:OUTPut[1]:SMODe <name>
Function
This Command is used to select the output-off mode
of the GSM.
With HIMPedance selected, the output relay opens
when the source is turned off. This disconnects
external circuitry from the GSM Input/ Output.
To prevent excessive wear on the output relay, do
not use the HIMPedance mode for tests that turn the
output on and off frequently.
With NORMal selected, the V-Source is selected and
set to 0V when the output is turned off. Compliance
is set to 0.5% full scale of the present current range.
In the ZERO output-off state when the V-Source
OUTPUT is turned off, the V-Source is set to 0V and
current compliance is not changed. When the I-
Source OUTPUT is turned off, the V-Source mode is
selected and set to 0V. Current compliance is set to
the programmed Source I value or to 0.5% full scale
of the present current range, which is greater.The