Software Control:
xT microscope Control Software
4-36
•
Switch off CCD automatically
(No / 1 minute / 10 minutes / 30 minutes / 1 hour / 2 hours / 6 hours)
Specifies if and when the CCD camera and infrared LEDs should
be automatically switched off. The countdown starts when
resuming the optical quad and continues regardless of the
operator activity.
•
Pause beam quads when switching off HV
(Yes / No)
Specifies if the electron image quads should be automatically
paused when switching off the High Voltage.
•
Allow Beam Shift in Get mode
(Yes / No)
Enables / disables automatic using of Beam Shift when the user
requires very small point-to-point movements (double-click in the
image at high magnifications).
•
Blank beam during long stage moves
(Yes / No)
If Yes is selected, the electron beam is automatically blanked
during long software controlled stage movements. This may
protect extremely sensitive samples from exposure to the beam in
undesired areas.
The Movie Tab
provides two groups of controls:
Timer
to set-up the movie frame-rate
and
File
to set-up the path name and format of the resulting movie
(see Chapter 5).
The Sensitivity Tab
The preset sliders control the sensitivity of the Manual User Interface
(
MUI
– option).
FIGURE 4-21
SENSITIVITY PREFERENCES
All MUI controls are represented except the Magnification. The
Default
button sets the original settings.
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...