Stages:
Stages Types and Accessories
7-6
STAGE MOVEMENT LIMITS
The motorised movements of the stage can be operated under
software control for more advanced location mapping. This includes
Shift
,
Get
,
Track
and the
Stage
module functionality. A live image can
be repositioned either by the stage movement (manual or software) or
by the
Beam Shift
. The tilting mechanism can be locked for more
stability at high magnifications using the software
Clamp
feature.
Note:
When moving the stage or tilting the specimen, the magnification may
need to be reduced not to lose the feature of interest off the screen.
C a u t i o n !
The positive Z value direction depends on the
Link Z to FWD
status
(see below).
FIGURE 7-6
STAGE MOVEMENT SCHEMA
TABLE 7-2
STAGE FEATURES AND LIMITS
Item
50 mm
100 mm
150 mm
Note
X
-25 to +25 mm -50 to +50 mm -75 to +75 mm
Y
-25 to +25 mm -50 to +50 mm -75 to +75 mm
Z
25 + 25 mm
0 + 60 mm
0 + 60 mm
in external
R
360°
360°
360°
continuous
T
-15° to +75°
-5° to +70°
-5° to +70°
Eucentric Position
WD = 10 mm
11.6 mm
*)
29.2
mm
*)
*)
distance between a sample
top surface and the stage base
Clamp
No Yes
Yes
Maximum sample
weight
250 g
2 000 g
5 000 g
for all tilt angles
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...