System Options:
Cooling Stage / Waterless Cooling Stage
9-23
FIGURE 9-20
COOLING STAGE ASSEMBLY
MOUNTING ADAPTER WITH MOUNTING SCREW
The Chamber feed-through Plate
provides feed-through connectors for the cables and water hoses.
FIGURE 9-21
CHAMBER FEED-THROUGH PLATE INSIDE / OUTSIDE
Cooling Stage Controller
This microprocessor-controlled board provides accurate and stable
automatic temperature control of the cooling stage and interfaces with
the xT microscope Control software.
The temperature measurement accuracy is determined by the
thermistor and the controller. These specifications are as follows:
•
thermistor accuracy:
± 0.1 °C
•
normal / limit temperature range:
± 20 °C from an ambient temperature / from - 25 to + 55 °C
Additionally, accuracy of a temperatures readout depends on sample
conductivity, thickness, shape and general thermoelectric properties
that vary by sample and mounting method.
Cables
•
The
Outer Cable
connects the outer feed-through plate side with
the controller board.
•
The
Inner Cable
connects the inner feed-through plate side with
the cooling stage (to which the cable is permanently attached).
Heating Stage connector
Water hoses flanges
(blue color)
Cooling Stage connector
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...