Stages:
Stages Types and Accessories
7-5
Sample Top Surface Positioning
The distance between the observed sample and the stage rotation
head surfaces must be properly set to bring the stage to the eucentric
position. This procedure also prevents the specimen to touch the lens
pole when moving the stage in the Z-axis direction.
The Z coordinate can now be changed via the manual or software Z
control around the eucentric position and further, but not less than
1 mm from the lens for safety reasons.
TABLE 7-1
SAMPLE POSITIONING PROCEDURE
50 mm
100 / 150 mm
Load a sample onto the specimen holder.
Adjust the external Z
coordinate as high as
possible.
Adjust the Z coordinate to bring the
specimen top surface
approximately 10 mm below the
lens.
Put the Eucentric Position
Adjuster on the stage base.
Bring the specimen top surface
point to the 2 mm adjuster
position by turning the
specimen holder internal screw.
Lock the position with the
locking cone.
Adjust the external Z
coordinate so that the
specimen top surface coincides
with the adjuster Eucentric
position.
Close the chamber and pump it down.
Switch on the beam, focus the specimen top surface and run the
Link Z to FWD function (see Chapter 4). The WD is now recognised
by the system as the
Stage
module /
Coordinates
tab Z value.
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...