
Software Control:
xT microscope Control Software
4-32
Slow
(tortoise / large green sector) /
Fast
(hare / small green sector)
preset icon indicates the matching dwell-time value. To change it
move an icon up or down just by pressing it.
Snapshot
(F4) /
Photo
(F2) (different cameras) preset icon indicates
the matching dwell-time value. Set all possible properties in the
Property editor.
The
Default
button restores the default dwell-time list and preset
settings.
The Presets Tab
allows the user to change the pre-set values in the
High Voltage
,
Magnification
and
Pressure
drop down lists. A value can be
changed by selecting it in the list and entering a new one in the edit
box just below the respective title. The new value replaces the
selected one and is immediately sorted into the list. The number of
entries in the list remains fixed.
FIGURE 4-18
PRESETS PREFERENCES
The
High Voltage
list can be changed to span any values from 200 V
to 30 kV. The values must be entered in kilovolts (0.2 means 200 V).
The
Magnification
list can be changed to hold frequently used
magnifications. Values that are in the pre-set list but cannot be applied
to the actual SEM conditions are not shown in the toolbar /
Magnification list box. Magnification range is from 20× to 1 000 000×.
Note:
Alternatively the Magnification could be displayed as the Horizontal
Field Width (see the
Preferences…
/
General
tab).
The
Pressure
list can be changed to hold specific values frequently
used in the Low Vacuum and ESEM modes in the range from 10 to
200 Pa (from 0.075 to 1.5 Torr).
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...