Alignments:
Quanta FEG System Alignments
6-3
BUTTONS AND CONTROL ELEMENTS
The following particular buttons and control elements have the same
behaviours for all alignment procedures, when available:
•
The
Start
button starts the procedure and proceeds with following
dialogues.
•
The
End
button moves the user to the last step (by clicking the
Next
button) to be able to finish the alignment procedure.
•
The
Contrast
/
Brightness
adjusters enable to optimize the image
quality during alignment.
•
The
Auto
button executes the appropriate alignment action
automatically for a particular voltage / spot / direction (whatever
suitable) with the use of the
Image Recognition
software. If this
utility does not recognize image features well, the procedure is
aborted and Warning message appears onscreen. In this case
change the imaging conditions (better focus, slower scanning, or
lower magnification) and try again.
•
The
Crossover
button activates the Crossover mode, where the
onscreen image shows the electron source tip instead of the
sample.