System Overview:
System Layout of Quanta FEG
2-6
Final Lens Aperture Strip
The strip is made from a Mo coated Si. Either manual (numbered
click-stop mechanism positions) or Automatic Aperture System (AAS)
motorized software-control (option - see Chapter 9) enables to choose
the aperture most applicable to your imaging needs (see Chapter 5).
The left hand turn of the large ring moves the aperture holder
outwards towards the larger aperture. After the aperture change, it is
necessary to tune its position by using the inner knob and the one on
the right side, controlling horizontal plane fine movements (see
Chapter 6).
FIGURE 2-6
FINAL LENS APERTURE STRIP CONTROL KNOB
WA R N I N G !
Aligning the final lens aperture is a mechanical process. It is
possible to misalign the aperture strip and obscure the beam
from reaching the sample. Adjustment should only be done by
those with the understanding of how to proceed. The aperture
holes edges cleanness is also very important (see Chapter 8).
TABLE 2-1
APERTURE SIZES AND THEIR USE
No. Standard
Option
Option
Recommended use
(related to Standard sizes)
1
1000 µm
1000 µm
1000 µm
Service Alignment (hole in frame)
2
-
-
-
3
50 µm
100 µm
30 µm
High current applications
4
40 µm
50 µm
30 µm
X-ray mapping of low-Z elements at low voltages
5
30 µm
40 µm
20 µm
General imaging or X-ray analysis
6
30 µm
30 µm
15 µm
Dynamic experiments
7
20 µm
30 µm
10 µm
High resolution imaging
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...