System Options:
Beam Deceleration
9-68
STAGE MODIFICATION
When extending the system capabilities with the Beam Deceleration
mode, it is necesary to make the stage modification by mounting the
BD attachment on the stage head. For systems acquired with the
Beam Deceleration this procedure is not carried out.
1.
Open the specimen chamber and remove the specimen holder.
2.
Place holding screw to the free stage position (use M3 or M4
thread according to which is avalilable).
3.
Place the BD attachment on the stage (holding screw is under the
cable output part to keep it stable while rotating the stage) and tight
the central screw to fix the attachment (use the socket-head
screw).
4.
Set maximum X,Y and Tilt coordinates to ensure the cable does
not limit stage movement. Put the cable into the holder, be sure the
cable is not placed in CCD camera field of view!
5.
Plug the connector from the BD attachment to the chamber
feedthrough and slightly rotate it to lock.
6.
Place the specimen holder back.
Be aware (before closing the chamber door) the specimen height
increases after mounting the BD attachment!
The BD attachment could be left in the chamber or removed, when
larger specimens should be observed.
Summary of Contents for Quanta FEG 250
Page 8: ...C vi...
Page 28: ...System Operation Quanta FEG System States 3 10...
Page 108: ...Alignments 1 Gun Alignment 6 6...
Page 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Page 152: ...System Options Automatic Aperture System 9 4...
Page 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Page 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...