Appendix D Optimizing the CANBERRA iSERIES for Air Sample Counting
The GAIN Calibration should only be necessary in the event of:
•
Change of Hardware
•
System Master Clear with no restorable personality file available
•
Radon compensation issues not resolved following the recommendations
included in this application note
When performing the gain calibration, the
i
SOLO (or
i
MATIC) automatically adjusts
the gain so the upper edge of the Am-241 peak falls in channel 136 using a successive
approximation routine. Figure 8 shows the spectrum from a typical Am-241 standard.
Figure 8 Typical Am-241 Spectrum