QC Data Display
QC Data Display
QC data is now available for display from within the View key menu. To access, from
the main screen press View. Select
#3 View QC Data
. All of the QC data selections
are shown in a menu:
QC Reports
1
α
Efficiency QC
2
β
Efficiency QC
3
α
Background QC
4
β
Background QC
Select the QC parameter desired and then press
Enter
. The system will display all QC
data points currently stored in the instrument. The following column information is
displayed for each data point:
View QC Data
Date Data Description Status
The date will reflect the date the data point was taken. The data description is
dependent on the type of data displayed. The status tells whether the data point is
within accepted limits or outside of the warning or lockout levels as defined by the QC
procedure.
iMatic Automatic Counting System User's Manual - 9239221K