Define QC Procedures
8.
Select the daily background lockout level. Default value is 0 (off). Press the
Enter
key to accept and automatically advance to the next question in the
Daily Bkg Parameters definition.
9.
Enter the carrier number of the source blank to be used to count the daily
background and automatically advance to the next question in the Define QC
Procedures definition.
Daily
α
Eff Parameters
The Daily
α
Eff Parameters menu allows you to define the daily
α
efficiency
measurement procedure.
To edit the daily a efficiency measurement procedure:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “10” in the menu,
Quality Control, and then press the
Enter
key.
3.
Using the arrow keys position the cursor “>” to item “3”, Daily
α
Eff Params,
and then press the
Enter
key.
4.
Using the arrow keys position the cursor “>” to item “1”, Count Time, and
then press the
Enter
key.
5.
Enter the daily
α
efficiency measurement time in minutes. Default value is
10
minutes
. Press the
Enter
key to accept and automatically advance to the next
question in the Daily
α
Eff Parameters definition.
6.
Enter the number of times the individual daily a efficiency measurement is to
be counted. Default value is
1
. Press the
Enter
key to accept and automatically
advance to the next question in the Daily
α
Eff Parameters definition.
7.
Select the daily
α
efficiency warning level. Default value is 0 (off). Press the
Enter
key to accept and automatically advance to the next question in the
Daily
α
Eff Parameters definition.
8.
Select the daily
α
efficiency lockout level. Default value is 0 (off). Press the
Enter
key to accept and automatically advance to the next question in the
Daily
α
Eff Parameters definition.
9.
Enter the carrier number of the source to be counted for the daily
α
efficiency
and automatically advance to the next question in the Define QC Procedures
definition.
iMatic Automatic Counting System User's Manual - 9239221K
63