Chapter 3 Setup Menu
Daily
β
Eff Parameters
The Daily
β
Eff Parameters menu item allows you to define the daily
β
efficiency
measurement procedure.
To edit the daily
β
efficiency measurement procedure:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “10” in the menu,
Quality Control, and then press the
Enter
key.
3.
Using the arrow keys position the cursor “>” to item “4”, Daily
β
Eff Params,
and then press the
Enter
key.
4.
Using the arrow keys position the cursor “>” to item “1”, Count Time, and
then press the
Enter
key.
5.
Enter the daily
β
efficiency measurement time in minutes. Default value is 10
minutes. Press the
Enter
key to accept and automatically advance to the next
question in the Daily
β
Eff Parameters definition.
6.
Enter the number of times the individual daily b efficiency measurement is to
be counted. Default value is 1. Press the
Enter
key to accept and automatically
advance to the next question in the Daily
β
Eff Parameters definition.
7.
Select the daily
β
efficiency warning level. Default value is 0 (off). Press the
Enter
key to accept and automatically advance to the next question in the
Daily
β
Eff Parameters definition.
8.
Select the daily
β
efficiency lockout level. Default value is 0 (off). Press the
Enter
key to accept and automatically advance to the next question in the
Daily
β
Eff Parameters definition.
9.
Enter the carrier number of the source to be counted for the daily
β
efficiency
and automatically advance to the next question in the Define QC Parameters
definition.
64
iMatic Automatic Counting System User's Manual - 9239221K