Define Report
Gross Counts
Alpha gross counts (3 to 5.6 MeV region), no background
subtraction
Net Counts
Alpha gross counts (3 to 5.6 MeV region) with background
subtracted
CPM
Alpha counts per minute (3 to 5.6 MeV region), no
background subtraction
Net CPM
Alpha counts per minute (3 to 5.6 MeV region) with
background subtracted
Activity X
±
Sigma
Total uncertainty associated with above alpha (3 to 5.6 MeV
region) Procedure Activity, Gross Counts, Net Counts,
CPM or Net CPM
Decision Level
Report decision level for above alpha (3 to 5.6 MeV region)
Procedure Activity, Gross Counts, Net Counts, CPM or Net
CPM
α
(3-5.0 MeV)
Procedure Activity
Alpha activity for region from 3 to 5.0 MeV is printed in the
unit (dpm, Bq,
µ
Ci, nCi, pCi or user defined K) selected in
the procedure used to perform the analysis
Gross Counts
Alpha gross counts (3 to 5.0 MeV region), no background
subtraction
Net Counts
Alpha gross counts (3 to 5.0 MeV region) with background
subtracted
CPM
Alpha counts per minute (3 to 5.0 MeV region), no
background subtraction
Net CPM
Alpha counts per minute (3 to 5.0 MeV region) with
background subtracted
Activity X
±
Sigma
Total uncertainty associated with above alpha (3 to 5.0 MeV
region) Procedure Activity, Gross Counts, Net Counts,
CPM or Net CPM
Decision Level
Report decision level for above alpha (3 to 5.0 MeV region)
Procedure Activity, Gross Counts, Net Counts, CPM or Net
CPM
Beta
Procedure Activity
Beta activity is printed in the unit (dpm, Bq,
µ
Ci, nCi, pCi or
user defined K) selected in the procedure used to perform
the analysis
Gross Counts
Beta gross counts, no background subtraction
Net Counts
Beta gross counts with background subtracted
CPM
Beta counts per minute, no background subtraction
Net CPM
Beta counts per minute with background subtracted
Activity X
±
Sigma
Total uncertainty associated with above beta Procedure
Activity, Gross Counts, Net Counts, CPM or Net CPM
Decision Level
Report decision level for above beta Procedure Activity,
Gross Counts, Net Counts, CPM or Net CPM
Limit
Action Limits, Warning and Alarm (defined in counting
procedure). See
Action Limits
on page 27.