Define Std Source
Source Calibration Month
To enter the month of calibration:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “7” in the menu, Define
Std Sources, and then press the
Enter
key.
3.
Using the arrow keys, position the cursor “>” to select the desired source to
edit, and then press the
Enter
key.
4.
Using the
Enter
key, advance to the Select Month prompt.
5.
Using the arrow keys or the numeric keypad, enter the month of calibration
(January through December). Press the
Enter
key to accept and advance to the
next question.
6.
If only the calibration month is modified, pressing the Cancel key will bring up
a “Save Changes?” question. Select
1. No
if the change was an error or
2.
Yes
to save the change(s) made.
Source Calibration Day
To enter the calibration day of the month:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “7” in the menu, Define
Std Sources, and then press the
Enter
key.
3.
Using the arrow keys, position the cursor “>” to select the desired source to
edit, and then press the
Enter
key.
4.
Using the
Enter
key, advance to the Enter Day prompt.
5.
Using the arrow keys or the numeric keypad, enter the calibration day of the
month (1 through 31). Press the
Enter
key to accept and advance to the next
question.
6.
If only the calibration day is modified, pressing the Cancel key will bring up a
“Save Changes?” question. Select
1. No
if the change was an error or
2.
Yes
to save the change(s) made.
iMatic Automatic Counting System User's Manual - 9239221K