Chapter 3 Setup Menu
Guard Det Serial Number
Guard Det Serial Number menu item allows you to enter the guard detector’s serial
number.
To enter a guard detector serial number:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “8” in the menu, Detector
Parameters, and then press the
Enter
key.
3.
Using the arrow keys position the cursor “>” to item “3”, Guard Det Serial #,
and then press the
Enter
key.
4.
Enter the serial number.
5.
Press the
Enter
key to accept and automatically advance to the next question
in the Detector Parameters definition.
6.
If you desire to only modify the Guard Det Serial number, pressing the
Cancel
key will return to the Detector Parameters Menu.
Sample Det HV
The Sample Det HV menu item allows you to set the sample detector high voltage from
0 V to 150 V. The unit will default to the last entered value.
Note: Changing the sample detector high voltage will cause the system to require a
complete calibration, starting with the detector and electronics gain calibration.
To enter a sample detector high voltage:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “8” in the menu, Detector
Parameters, and then press the
Enter
key.
3.
Using the arrow keys position the cursor “>” to item “4”, Sample Det HV, and
then press the
Enter
key.
4.
Enter the Sample Det HV.
5.
Press the
Enter
key to accept. If the value has been changed, a warning
message will be displayed:
52
iMatic Automatic Counting System User's Manual - 9239221K