Define Procedure
6.
Press the
Enter
key to accept and automatically advance to the next parameter
in the procedure. Press the
Cancel
key to return to the Define Procedure
Menu. Pressing the
Cancel
key again will bring up a “Save Changes?”
question. Select
1. No
if the change was an error or
2. Yes
to save the
change(s) made.
Count Presets
The Count Presets menu item allows you to enter the maximum number of Alpha or
Beta counts to terminate data collection is set by the Count Presets parameters. You
can enter a value from 1 - 99,999,999; defaults to 99,999,999.
To enter the Count Preset:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “3” in the menu, Define
Procedure, and then press the
Enter
key.
3.
Using the arrow keys, select the procedure, and then press the
Enter
key.
4.
Using the arrow keys, position the cursor “>” to item “7”, Count Presets, and
then press the
Enter
key.
5.
Enter the maximum number of Alpha or Beta counts to terminate data
collection.
6.
Press the
Enter
key to accept and automatically advance to the next parameter
in the procedure. Press the
Cancel
to return to the Define Procedure Menu.
Pressing the
Cancel
key again will bring up a “Save Changes?” question.
Select
1. No
if the change was an error or
2. Yes
to save the change(s)
made.
iMatic Automatic Counting System User's Manual - 9239221K