Chapter 3 Setup Menu
Beta Comp
Procedure Activity
Radon compensated beta activity is printed in the unit (dpm,
Bq,
µ
Ci, nCi, pCi or user defined K) selected in the
procedure used to perform the analysis.
Activity X
±
Sigma
Total uncertainty associated with above Radon compensated
beta Procedure Activity
Decision Level
Report decision level for above Radon compensated beta
Procedure Activity.
Limit
Action Limits, Warning and Alarm (defined in counting
procedure). See
Action Limits
on page 27.
Radon
Contribution
Procedure Activity
Radon background region activity is printed in the unit
(dpm, Bq,
µ
Ci, nCi, pCi or user defined K) selected in the
procedure used to perform the analysis
Gross Counts
Radon background region gross counts, no background
subtraction
CPM
Radon background region counts per minute, no
background subtraction
Activity X
±
Sigma
Total uncertainty associated with above Radon background
region Procedure Activity, Gross Counts, or CPM
Decision Level
Report decision level for above Radon background region
Procedure Activity, Gross Counts, or CPM
Thoron
Contribution
Procedure Activity
Thoron background region activity is printed in the unit
(dpm, Bq,
µ
Ci, nCi, pCi or user defined K) selected in the
procedure used to perform the analysis
Gross Counts
Thoron background region gross counts, no background
subtraction
CPM
Thoron background region counts per minute, no
background subtraction
Activity X
±
Sigma
Total uncertainty associated with above Thoron background
region Procedure Activity, Gross Count, or CPM
Decision Level
Report decision level for above Thoron background region
Procedure Activity, Gross Counts, or CPM
Guard
Gross Counts
Gross counts from guard detector
Gross Rate
Count rate (cpm) from guard detector
40
iMatic Automatic Counting System User's Manual - 9239221K