Define Procedure
If you choose to use user-defined conversion factor (K User Defined):
1.
Select
User
Defined Unit.
2.
Enter the custom conversion factor value from 0.00 to 1.00 x 10
18
.
3.
Enter the custom conversion factor sigma from 0.00 to 1.00 x 10
18
.
4.
Enter the custom conversion factor label up to 8 alphanumeric characters.
5.
Press the
Enter
key to accept and automatically advance to the next question
in the procedure definition.
6.
If only the Conversion Factor and/or Sigma are modified, pressing the
Cancel
key will return to the Define Procedure Menu. Pressing the
Cancel
key again
will bring up a “Save Changes?” question. Select
1. No
if the change was an
error or
2. Yes
to save the change(s) made.
Action Limits
The Action Limits are user-defined levels that are compared to the sample counts at the
completion of the measurement and are used to create a report action flag.
The iMatic allows four individual alpha and beta action limits to be independently set
for each user procedure (Procedures A - J). Two additional lines in the report header
area print the procedure action limit values.
The action limit values can also be printed on the report by selecting Alpha Limit or
Beta Limit as a field that is defined in the report. The sample warning flags are printed
in a Limit column as
*WARNING*
if the Warning level is exceeded and as
*ALARM*
if the Alarm Limit is exceeded.
Action Limits set the individual warning and alarm level limits for alpha and beta
counts in the same units selected as the Conversion Factor. You can enter a value from
0.00 to 1.0 x 10
18
.
To enter the Action Limits:
1.
Press the
Setup
key.
2.
Using the arrow keys, position the cursor “>” to item “3” in the menu, Define
Procedure, and then press the
Enter
key.
3.
Using the arrow keys, select the procedure, and then press the
Enter
key.
4.
Using the arrow keys, position the cursor “>” to item “9”, Action Limits, and
then press the
Enter
key.
iMatic Automatic Counting System User's Manual - 9239221K
27