773232AH
4-2
Specifications
Instrument Specifications And Characteristics
Cytometer
Flow Cell
Sensing area: BioSense 150
μ
m x 460 rectangular channel with an integral lens, mounted with a
vertical (upward) flow path. See
.
Flow Rate
Continuous pressure is applied to the sample tube. The amount of pressure depends on the flow rate
you specify:
•
Low approximately 10 uL/min
•
Medium approximately 30 uL/min
•
High approximately 60 uL/min
Sheath Consumption
•
780mL / hour (acquisition)
•
0mL / hour (standby)
Lasers
•
Solid-state, software controlled, 22 mW, blue laser operating at 488 nm and
•
Solid-state, software controlled, 25 mW, red laser operating at 638 nm.
Optional third laser:
•
Solid-state, software controlled, 40 mW, violet laser operating at 405 nm.
Laser Power Monitoring
Laser power is monitored by each laser individually. If the laser power deviated more than ±10% a
Laser Power Error is displayed on the
. The system will not run a sample until corrective
measures rectify the fault. Follow the instructions in the Troubleshooting section for handling this
error.
Beam-Shaping Optics
Cross cylindrical lenses 10 mm by 80 mm.
Blue Laser Beam Spot Size
An elliptical spot 10
μ
m high by 84
μ
m wide.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......