773232AH
8-2
Running Samples
Running Samples - MCL Automatic Mode
CAUTION
Risk of erroneous results if the Cytometer has been idle for an extended period of
time or you have just performed Daily Startup. To ensure correct results, perform
a prime after:
• Daily Startup.
• The Cytometer has been idle for an extended period of time.
• You place a new carousel on the MCL and light scatter signals appear
abnormal.
CAUTION
Risk of sample misidentification if a power failure occurs during sample
processing. In the event of a power failure, discard any in-process samples.
Running Samples - MCL Automatic Mode
Verifying A Worklist
Always verify carousel and tube location fields before starting acquisition and ensure that the
Acquisition Manager information accurately reflects the tubes in the carousel being run. The order
of the tubes in the carousel must match the order of the tubes in the Worklist.
The default protocol (Ctrl N) must be saved prior to being used in the Worklist. If the default
protocol is not saved with a specific name prior to using it, that tube may be run with unexpected
protocol settings depending on the Save Protocol selection in Workspace Preferences. Also, the
default protocol name will default to "setup.def" in any associated panel template and have no
regions.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......