773232AH
13-19
Troubleshooting
Cytometer Messages Table
13
Front Panel Connection Fault – SCA2
A fault was detected in the
connection between the data
acquisition board SCA 2 and the front
panel display board.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Front Panel Connection Fault – SCA3
A fault was detected in the
connection between the data
acquisition board SCA 3 and the front
panel display board.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Hardware Made Safe
An error has occurred or a request
has been made from CytoTools and
the harware has been put into a safe
state.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Host Command Invalid
The Cytometer software could not
process the requested command
from the Workstation.
1.
Reboot the computer and restart the
Cytometer.
2.
If problem continues, contact your local
Beckman Coulter Representative.
Incorrect Laser Type - Blue
The type of the blue laser does not
agree with the type in the instrument
configuration.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Incorrect Laser Type - Red
The type of the red laser does not
agree with the type in the instrument
configuration.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Incorrect Laser Type - Violet
The type of the violet laser does not
agree with the type in the instrument
configuration.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Internal Sheath Tank Level Error
There is no more sheath fluid in the
internal sheath tank and the
acquisition has been stopped.
If error persists, contact your local Beckman
Coulter Representative.
Internal Sheath Tank Level Warning
The internal sheath tank is empty.
If error persists, contact your local Beckman
Coulter Representative.
MCL Door Open Error
MCL cover is open while the MCL is in
use.
Close the MCL cover.
MCL Door Open Warning
MCL cover is open while the MCL is in
use.
Close the MCL cover.
Optics Board Not Found
The optics I/O board could not be
detected.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Optics I/O Board Voltage Error
The voltages on the optics I/O board
are out of range.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......